X-ray aparatuses RAP 50-RAP 200

Main
Characteristics
Documents
Radiographs

X-ray apparatus family of RAP are designed for a independent problem solving industrial radiographic and special control, medical diagnostics and research.

 

Components of X-ray apparatus RAP family:

  • X-ray source (RI series);
  • Remote control on the basis of microprocessor devices.

With the help of the remote control is performed on and off the device, selection, installation and indication of modes of operation, the output of the self-diagnosis results. It is also possible to control the work of the apparatus from an external computer.

 

Apparatus of «RAP» series have:

  • high operational reliability
  • stability of parameters throughout the operating range;
  • the ability to configure the operating modes for specific tasks.

The main technical characteristics of the apparatus «RAP» series:

Parameter Value
Implementation of X-ray apparatus monoblock, cable, portable monoblock
Power supply of X-ray machine 1. ~ 220V/50Hz (AC)
2. – 48 V (DC)
3. Accumulator
Power consumption up to 0,5 kV
Anode voltage range depending on the application of device 20 - 200 kV
Step adjustment anode voltage depending on the maximum operating voltage 1kV, 5kV, 10kV
The maximum deviation of the anode voltage depending on the application of device 0.1%, 0.5%, 1%
The anode current range depending on the application of the systems 0,02 - 5 mA
Step adjustment of the anode current depending on the anode voltage 0.005 mA, 0.01 mA, 0.1 mA
The maximum deviation of the anode current, depending on the purpose of the systems 0.5%, 1%, 5%
The range of focal spot sizes depending on the application of the systems 0.03х0.03 mm, 0.05х0.05 mm, 0.1х0.1 mm, 0.5х0.5 mm
Modes of operation intermittent, longtime, continuous
The time to working mode depending on the application of the systems 0.05 - 10 sec
The range of powers on the anode of X-ray tube, depending on the purpose of the systems 2 - 150 W
Control of the power, the operating mode and the exposure time  existence 
Variants of the target of X-ray tube  massive side, a massive panoramic, transmitting 
X-ray exit angle  directed - 40°х60°,30°х30°, panoramic - 160°  
Filtering of X-rays  the equivalent of Al - 2mm 
Variants of the cooling system air, forced air, water
High-voltage insulation is X-ray source solid-state, oily, combination
Management of parameters of X-ray source manual, software, remote, the ability to sync with visualization systems and actuators
Communication with the computer RS-232
Possible X-ray system of registration X-ray film, system with fotostimulated luminophor, semiconductor sensors, electronic panel